UltiMate 3000 Series:
VWD-3100 and VWD-3400RS Variable Wavelength Detectors
Operating Instructions Page 13
1.3 Intended Use
For Research Use Only. Not for use in diagnostic procedures. The device is designed to be
operated only be qualified and authorized personnel. All users must know the hazards
presented by the device and the used substances.
The detector is designed for laboratory research use in high-performance liquid
chromatography (HPLC) applications. It is part of the UltiMate 3000 system, but can also
be used with other HPLC systems if adequate control inputs and outputs are available. A
PC with USB port is required.
The detector can be operated either directly from the front panel display or with the
Chromeleon Chromatography Management System. Being part of the UltiMate 3000 system,
the detector can also be operated with other data systems, such as
Xcalibur™, Compass™/HyStar™, or Analyst
®
.
Installation of the DCMS
Link
(Thermo Scientific Dionex Chromatography Mass
Spectrometry Link) software is required in addition to the installation of the data system.
Empower™.
Installation of the Thermo Scientific Dionex Instrument Integration Software is required
in addition to the installation of the data system.
For more information, contact the Thermo Fisher Scientific sales organization for Dionex
HPLC Products.
Note the following:
Note that the detector may be operated only with accessories and spare parts
recommended by Thermo Fisher Scientific ( page 139) and within its technical
specifications ( page 135).
Use only standard solvents of at least HPLC grade or better LC-MS grade (0.2 µm,
filtered), and buffers that are compatible with the flow path materials. Note the special
properties of the solvents, such as the viscosity, boiling point, and UV absorption
(UV/VIS detector).
Buffer concentration: Typically up to 1 mol/L (stainless steel flow cells: < 0.1 mol/L
chloride ions).
Also observe the information about the solvent compatibility and buffer concentrations of
the other UltiMate 3000 system modules. For more information, refer to the Operating
Instructions for the modules.